<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>40345</ErrorNumber>
<ErrorFamily>Media Handler</ErrorFamily>
<ErrorMessage>
MDH: Please remove plate(s) from the outer bay [&lt;I/O name&gt;]
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Trendsetter_400/800</Text></ListItem>
<ListItem><Text>Trendsetter_400/800_Autoloader</Text></ListItem>
<ListItem><Text>Trendsetter_NEWS</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
A plate has been detected on the outer bay, and must be removed.  This could occur during
initialization.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Remove the plate(s) from the outer bay.  This may involve undocking the preloader.
</text></DescriptionText>
</OperatorAction>
<Availability>
<DescriptionText><text>
This error is reported.
</text></DescriptionText>
</Availability>
<Cause>
Plate(s) left on outer bay when machine was reset or workstation disconnected
</Cause>
<Parameter>
I/O name -
This is the standard firmware name of the sensor that indicates a plate on the outer bay.  It may
also say "stored state", which means that the firmware has an internal flag to track whether or
not there is media on the outer bay.
</Parameter>
<LastModified>
11-May-2007
</LastModified>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 1 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 1 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllParameters.htm</LinkAddress><LinkTitle>Firmware 1 Parameter Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Thu Jul 17 07:23:25 2008 PST
</Updated>
<Links>
</Links>
</Error>